The Vibration Controllers exploit the power of new generation digital signal processing techniques to deliver very high degree of control accuracy and performance in vibration test applications.

The 16/32 Bit Fixed/Floating point processors specifically designed to process complex mathematical calculations & algorithms, combined with high speed simultaneous data sampling Analog-to-Digital Converters and Digital-to-Analog Converters make the whole architecture best suitable for developing all types of vibration control applications to fulfill everyday demands of industry standards, research, Space and MIL applications worldwide.
Sine Vibration Controller
    Swept Sine Software  Specifications
Parameters
Model - I
Model - II
Model - III
Frequency Range 1 Hz – 5000 Hz 0.1 Hz – 5000 Hz 0.08 Hz – 5000 Hz
Control Parameter Constant – Acc./Vel/Disp
Slope - Acceleration
Constant – Acc./Vel/Disp..
Slope - Acc.//Vel/Disp
Constant – Acc./Vel/Disp..
Slope - Acc.//Vel/Disp
Test Mode Sweep Frequency Test
Fixed Frequency Test
(Discrete Step Control)
Resonance Search & Dwell
Manual Control using Keyboard
DVA Monitoring
(using ext. sinusoidal frequency input)
Sweep Frequency Test
Fixed Frequency Test
(Discrete Step Control)Resonance Search & Dwell (Optional)
-
-
Sweep Frequency Test
Fixed Frequency Test
(Discrete Step Control)
Resonance Search & Dwell (Optional)
-
-
Input Channels 02/04 08/16 32, expandable upto 128
Channel sensitivity 1.0 to 99.99 pC/g or mV/g 1.0 to 1000 pC/g or mV/g 0.1 to 10000 pC/g or mV/g
Channel Control Single/ Multi-channel Single/ Multi-channel Single/ Multi-channel
Control Mode
Filters
Maximal/ Average Maximal/ Average/ Weighted Average Maximal/ Average/ Weighted Average
8th Order (analog) Anti-aliasing (Analog/Digital) Anti-aliasing (Analog/Digital/Tracking)
Output Channels
Filters
Drive - 12V(Pk-Pk)
COLA- proportional to frequency
Drive - 20V(Pk-Pk)
COLA- proportional to frequency
Drive - 20V(Pk-Pk)
COLA- proportional to frequency
8th Order (analog) Anti-aliasing (Analog/Digital) Anti-aliasing (Analog/Digital/Tracking)
Digital I/O 4 I/P, 4 O/P 8 I/P, 8 O/P 16 I/P, 16 O/P
Compression Rate 0.001 to 3.0 dB/Sec/Hz 0.001 to 20 dB/Sec/Hz 0.001 to 20 dB/Sec/Hz
Compression Limit 2.0 to 200.0 dB/s
Control Dynamic Range > 85 dB > 95 dB >105 dB
Test Duration Upto specified no. of sweeps or infinite.
Upto specified duration of time.
Upto specified no. of sweeps.
Upto specified duration of time.
Upto specified no. of sweeps.
Upto specified duration of time.
Online Display Parameters Parameters Parameters
  Frequency, Programmed Control
Level, Measured Control Level,
Servo Voltage, No. of sweeps
elapsed/ remaining, Date , Time,
Displacement/Velocity/Acceleration
of each channel.

Frequency, Programmed Control
Level, Measured Control Level,
Servo Voltage, No. of sweeps
elapsed/ remaining, Date , Time,
Displacement/Velocity/Acceleration
of each channel.

Frequency, Programmed Control
Level, Measured Control Level,
Servo Voltage, No. of sweeps
elapsed/ remaining, Date , Time,
Displacement/Velocity/Acceleration
of each channel.
  Graph Graph Graph
  Control Curve (Control Vs.
Frequency)
Drive Curve (Drive Vs. Frequency)
Individual Channel Response
(Channel Values Vs. Frequency)
Transfer Function (Channel ‘Y’/
Channel ‘X’ Vs. Frequency)
Alarm/ Abort limit region
Control Curve (Control Vs.
Frequency)
Drive Curve (Drive Vs. Frequency)
Individual Channel Response
(Channel Values Vs. Frequency)
Transfer Function (Channel ‘Y’/
Channel ‘X’ Vs. Frequency)
Alarm/ Abort limit region
Composite Display
Control Curve (Control Vs.
Frequency)
Drive Curve (Drive Vs. Frequency)
Individual Channel Response
(Channel Values Vs. Frequency)
Transfer Function (Channel ‘Y’/
Channel ‘X’ Vs. Frequency)
Alarm/ Abort limit region
Composite Display
Graph
Annotations
Selection of Y-axis value as
acceleration/Velocity/Displacement.
Two lines of annotation programmable
by the operator for each test to identify
the test conducted.
Selection of Y-axis value as
acceleration/Velocity/Displacement.
Two lines of annotation programmable
by the operator for each test to identify
the test conducted.
Selection of Y-axis value as
acceleration/Velocity/Displacement.
Two lines of annotation programmable
by the operator for each test to identify
the test conducted.
Storage of
Test Data
To disk at pre-configured intervals in
tabular text format.
Export of test data to Microsoft Excel.
To disk at pre-configured intervals in
tabular text format.
Export of test data to Microsoft Excel.
To disk at pre-configured intervals in
tabular text format.
Export of test data to Microsoft Excel.
Storage of
Test Graph
Storage of Graph in WMF/ BMP format
on mouse click.
Export of test graph to Microsoft Word.
Storage of Graph in WMF/ BMP format
on mouse click.
Export of test graph to Microsoft Word.
Storage of Graph in WMF/ BMP format
on mouse click.
Export of test graph to Microsoft Word.
Posttest Report
Generation
Facilitates report generation of test
conducted.
Export of graph to Microsoft Word.
Facilitates report generation of test
conducted.
Export of graph to Microsoft Word.
Facilitates report generation of test
conducted.
Export of graph to Microsoft Word.
Safety features/
Protections
Loop check feature to verify physical
setting of the system before staring
actual test.
Independent alarm and abort profile in
terms of + (0.5 to 20 dB) of reference
test profile over the entire frequency
range.
Auto test shutdown in case of
exceeding the maximum ‘g’ level
programmed.
Auto test shutdown in case of control
exceeding the abort limit.
Password authorization to operate the
software.
Loop check feature to verify physical
setting of the system before staring
actual test.
Independent alarm and abort profile in
terms of + (0.5 to 20 dB) of reference
test profile over the entire frequency
range.
Auto test shutdown in case of
exceeding the maximum ‘g’ level
programmed.
Auto test shutdown in case of control
exceeding the abort limit.
Password authorization to operate the
software.
Loop check feature to verify physical
setting of the system before staring
actual test.
Independent alarm and abort profile in
terms of + (0.5 to 20 dB) of reference
test profile over the entire frequency
range.
Auto test shutdown in case of
exceeding the maximum ‘g’ level
programmed.
Auto test shutdown in case of control
exceeding the abort limit.
Password authorization to operate the
software.
       

   Broadband Random Control Software
Parameters Model - I Model - II Model - III
Test Mode Pseudo Random
True Random
True Random True Random
Frequency Range 500, 1000, 2000, 3000 or 4000 Hz 125,250,500,1000,2000 or 4000 Hz 31.25,62.5,125,250,500,1000,2000 or 4000 Hz
Frequency Lines Frequency Lines 100, 200, 300,400 or 500 100, 200, 400 or 800(1600 optional)
Input Channels 02/04 08/16 32, expandable upto 128
Channel type Selectable for Charge/ mV/ ICP Selectable for Charge/ mV/ ICP Selectable for Charge/ mV/ ICP
Channel sensitivity 1.0 to 99.99 pC/g or mV/g 1.0 to 1000 pC/g or mV/g 0.1 to 10000 pC/g or mV/g
Channel Control Single/ Multi-channel Single/ Multi-channel Single/ Multi-channel
Control Mode
Filters
Maximal/ Average Maximal/ Average/ Weighted Average Maximal/ Average/ Weighted Average
8th Order (analog) Anti-aliasing (Analog/Digital) Anti-aliasing (Analog/Digital/Tracking)
Output Channels
Filters
Drive - 20V (Peak-Peak) Drive - 20V (Peak-Peak) Drive - 20V (Peak-Peak)
8th Order (analog) Anti-aliasing (Analog/Digital) Anti-aliasing (Analog/Digital/Tracking)
Drive Signal Continuous Gaussian Random Continuous Gaussian Random Continuous Gaussian Random
Dynamic Range > 82 dB > 92 dB > 102 dB
Limit Channel Upto 200 Grms Upto 200 Grms Upto 200 Grms
Correction Rate 0.1 to 50.0 dB/sec 0.1 to 50.0 dB/sec 0.1 to 50.0 dB/sec
Crest Factor/ Sigma Clipping 1.8 to 8 2 to 8 2 to 8
Degree of Freedom 5 to 800 2 to 1000 8 to 8000
Loop time (typical) 200 mS

100 mS

100 mS
Test Duration Upto 499:59:59 Hrs Upto 999:59:59 Hrs Upto 9999:59:59 Hrs
Test Profile/ Spectrum Programmable thru tabular editor.
100 segments/ break points.
Simple editing of dB/Oct or PSD/ ASD
values
Insertion, deletion & changes of levels
& slope.
Capable to virtually programme any test
profile.
Graphical presentation of profile for
visualization.
Programmable thru tabular editor.
200 segments/ break points.
Simple editing of dB/Oct or PSD/
ASD values
Insertion, deletion & changes of
levels & slope.
Capable to virtually programme any
test profile.
Graphical presentation of profile for
visualization.
Programmable thru tabular editor.
200 segments/ break points.
Simple editing of dB/Oct or PSD/ ASD
values
Insertion, deletion & changes of levels &
slope
Capable to virtually programme any test
profile.
Graphical presentation of profile for
visualization.
Soft Test Startup Time 1 to 30 seconds 1 to 30 seconds 1 to 30 seconds
Soft Test Stop Time 1 to 30 seconds 1 to 30 seconds 1 to 30 seconds
Pre-Test Loop to determine
system response before
actually starting the main test
User selectable drive level from 5 to
2000 mV rms
User selectable drive level from 0 to
1000 mV rms
User selectable drive level from 0 to 1000
mV rms
Test Level Scheduling User selectable : -27 to 0 dB
Duration for each level can be
programmed separately.
User selectable : -80 to 0 dB
Duration for each level can be
programmed separately.
User selectable : -80 to 0 dB
Duration for each level can
programmed separately.
Online Display Parameters:
Programmed Grms, Control Grms,
Individual Channel(s) Grms, Channel(s)
in Use, Scheduled Test Level, Elapsed
Test Period, Remaining Test Period,
Drive Voltage (rms), Time & Date.

Graph:
Control/ Reference test spectra (g2/Hz
Vs. Frequency)
Individual channel spectra (g2/Hz Vs.
Frequency)
Overlaid Spectra ((g2/Hz Vs.
Frequency)
Transfer function spectra CHx/CHy (
dB Vs. Frequency)
Drive spectra ( dB Vs. Frequency)
Drive signal ( Volt Vs. Frequency)
Alarm/ Abort limits in all spectra.
Display Format:
Log or linear scale.
Alarm & Abort line display.
X & Y-axis grid enable/ disable
Parameters:
Programmed Grms, Control Grms,
Individual Channel(s) Grms,
Channel(s) in Use, Scheduled Test
Level, Elapsed Test Period,
Remaining Test Period, Drive
Voltage (rms), Time & Date.

Graph:
Control/ Reference test spectra (g2/Hz
Vs. Frequency)
Individual channel spectra (g2/Hz Vs.
Frequency)
Overlaid Spectra ((g2/Hz Vs.
Frequency)
Transfer function spectra CHx/CHy (
dB Vs. Frequency)
Drive spectra ( dB Vs. Frequency)
Drive signal ( Volt Vs. Frequency)
Alarm/ Abort limits in all spectra.
Display Format:
Log or linear scale.
Alarm & Abort line display.
X & Y-axis grid enable/ disable
Parameters:
Programmed Grms, Control Grms,
Individual Channel(s) Grms, Channel(s) in
Use, Scheduled Test Level, Elapsed Test
Period, Remaining Test Period, Drive
Voltage (rms), Time & Date.

Graph:
Control/ Reference test spectra (g2/Hz
Vs. Frequency)
Individual channel spectra (g2/Hz Vs.
Frequency)
Overlaid Spectra ((g2/Hz Vs.
Frequency)
Transfer function spectra CHx/CHy (
dB Vs. Frequency)
Drive spectra ( dB Vs. Frequency)
Drive signal ( Volt Vs. Frequency)
Alarm/ Abort limits in all spectra.
Display Format:
Log or linear scale.
Alarm & Abort line display.
X & Y-axis grid enable/ disable
Graph Annotations to
identify the test conducted
Two lines of annotation programmable
by the operator for each test
User configurable message/ notes User configurable message/ notes
Storage of Test Data To disk at pre-configured intervals in
tabular text format.
Export of test data to Microsoft Office
To disk at pre-configured intervals in
tabular text format.
Export of test data to Microsoft Office
To disk at pre-configured intervals in
tabular text format.
Export of test data to Microsoft Office
Posttest Report Generation Tabular data for PSD Vs. Frequency
Control/ Reference spectrum
Control test spectra (g2/Hz Vs.
Frequency)
Individual channel spectra (g2/Hz Vs.
Frequency)
Overlaid Spectra ((g2/Hz Vs.
Frequency)
Transfer function spectra CHx/CHy (
dB Vs. Frequency)
Drive spectra ( dB Vs. Frequency)
Tabular data for PSD Vs. Frequency
Control/ Reference spectrum
Control test spectra (g2/Hz Vs.
Frequency)
Individual channel spectra (g2/Hz Vs.
Frequency)
Overlaid Spectra ((g2/Hz Vs.
Frequency)
Transfer function spectra CHx/CHy (
dB Vs. Frequency)
Drive spectra ( dB Vs. Frequency)
Tabular data for PSD Vs. Frequency
Control/ Reference spectrum
Control test spectra (g2/Hz Vs.
Frequency)
Individual channel spectra (g2/Hz Vs.
Frequency)
Overlaid Spectra ((g2/Hz Vs.
Frequency)
Transfer function spectra CHx/CHy (
dB Vs. Frequency)
Drive spectra ( dB Vs. Frequency)
Safety features Loop check for proper physical
connections.
Check of programming values
limits.
Password authorization for secured
operation.
Max. ‘Grms' limit for each channel.
No. of lines for line abort
No. of loops for line abort
Loop check for proper physical
connections.
Check of programming values within
limits.
Password authorization for secured
operation.
Max. ‘Grms' limit for each channel.
No. of lines for line abort
No. of loops for line abort
Loop check for proper physical
connections.
Check of programming values within
limits.
Password authorization for secured
operation.
Max. ‘Grms' limit for each channel.
No. of lines for line abort
No. of loops for line abort
 
  Classical Shock Controller/Arbitrary Transient / SRS Analysis & Synthesis
Parameters Model - I Model - II Model - III
Modes of Operation Half Sine
Triangular
Forward Saw-tooth
Backward Saw-tooth
Rectangular
Half Sine
Triangular
Forward Saw-tooth
Backward Saw-tooth
Rectangular
Trapezoidal
Arbitrary Transient
SRS Analysis & Synthesis
Half Sine
Triangular
Forward Saw-tooth
Backward Saw-tooth
Rectangular
Trapezoidal
User defined shapes
Arbitrary Transient
SRS Analysis & Synthesis
Pulse amplitude 1 to 200 ‘g' 0.1 to 200 ‘g' 0.1 to 500 ‘g'
Pulse width 0.5 to 50 mS 0.1 to 100 mS 0.1 to 1000 mS
Pre & Post Pulse Amplitude
( % of Pulse Amplitude
5.0% to 50.0%(sine) 5.0% to 50.0%(sine, rectangular) 5.0% to 50.0%(sine, rectangular)
Input Channels 02/04 08/16 32, expandable upto 128
Channel sensitivity 1.0 to 99.99 pC/g or mV/g 1.0 to 1000 pC/g or mV/g 0.1 to 10000 pC/g or mV/g
Channel Control Single/ Multi-channel Single/ Multi-channel Single/ Multi-channel
Control Mode
Maximal/ Average Maximal/ Average/ Weighted Average Maximal/ Average/ Weighted Average
Filters 8th Order (analog) Anti-aliasing (Analog/Digital) Anti-aliasing (Analog/Digital/Tracking)
Control Mode Average Average/Maximal/Weighted Average/Maximal/Weighted
Channel Selection Disable/Monitor /Control & Monitor Disable/Monitor /Control & Monitor Disable/Monitor /Control & Monitor
Output Channel 8th Order (Analog) Anti-aliasing (Analog/Digital) Anti-aliasing (Analog/Digital/Tracking)
Filters 4 I/P, 4 O/P 8 I/P, 8 O/P 16 I/P, 16 O/P
Pre-Test Amplitude
(% of Reference Amplitude)
10.0% to 50.0%
Pre-test is carried out to determine
response/ transfer function of the
Vibration System by sending a low
level signal. This feature expedites the
equalization process at the time of
actual testing and safeguards the
product from getting over tested due
to pulses consumed in equalization.
5% to 50.0%
Pre-test is carried out to determine
response/ transfer function of the
Vibration System by sending a low
level signal. This feature expedites
the equalization process at the time
of actual testing and safeguards the
product from getting over tested due
to pulses consumed in equalization
5% to 50.0%
Pre-test is carried out to determine
response/ transfer function of the
Vibration System by sending a low
level signal. This feature expedites the
equalization process at the time of
actual testing and safeguards the
product from getting over tested due
to pulses consumed in equalization
Pulse Repetition Rate 50.0 to 3000 mS 20.0 to 3000 mS 10.0 to 5000 mS
Total no. of pulses to be
generated
1 to 100000 1 to 100000 1 to 100000
Online Display Parameters:
Pulse Width, Reference Pulse
Amplitude (g), Control Pulse
Amplitude (g), Drive Signal (V), Total
Pulse, Pulse Elapsed, Pulse
Remaining, Individual Channel Level,
Channel Mode (Control/ Monitor),
Displacement (mm), Velocity (mm/s),
Date, Time.

Graph:
Control graph (g vs. time), Individual
channel graph (g vs. time), Drive
graph (volt vs. time), Alarm/ Abort
limit region
Parameters:
Pulse Width, Reference Pulse
Amplitude (g), Control Pulse
Amplitude (g), Drive Signal (V), Total
Pulse, Pulse Elapsed, Pulse
Remaining, Individual Channel Level,
Channel Mode (Control/ Monitor),
Displacement (mm), Velocity (mm/
s), Date, Time.

Graph:
Control graph (g vs. time), Individual
channel graph (g vs. time), Drive
graph (volt vs. time), Alarm/ Abort
limit region
Parameters:
Pulse Width, Reference Pulse
Amplitude (g), Control Pulse
Amplitude (g), Drive Signal (V), Total
Pulse, Pulse Elapsed, Pulse
Remaining, Individual Channel Level,
Channel Mode (Control/ Monitor),
Displacement (mm), Velocity (mm/s),
Date, Time.

Graph:
Control graph (g vs. time), Individual
channel graph (g vs. time), Drive
graph (volt vs. time), Alarm/ Abort
limit region
Graph Annotations Two lines of annotation
programmable by the operator for
each test to identify the test
conducted.

User configurable message/ notes

User configurable message/ notes
Storage of Test Data To disk at pre-configured intervals in
tabular text format & waveforms
To disk at pre-configured intervals in
tabular text format & waveforms
To disk at pre-configured intervals
in tabular text format & waveforms
Storage of Test Graph Storage of waveform into user
defined file.
Export of test graph to Microsoft
Office
Storage of waveform into user
defined file.
Export of test graph to Microsoft
Office
Storage of waveform into user
defined file.
Export of test graph to Microsoft
Office
Posttest Report Generation Facilitates report generation of test
conducted.
Export of graph to Microsoft Office
Facilitates report generation of test
conducted.
Export of graph to Microsoft Office
Facilitates report generation of test
conducted.
Export of graph to Microsoft Office
Safety features/
Protections
Loop check feature thru Pre-test to
verify physical setting of the system
before starting actual test.
Independent alarm and abort profile
in terms of ± (1.0 to 50.0 %) of
reference pulse acceleration.
Auto test shutdown in case of
control exceeding the abort limit or
lines exceeds for set no. of loops.
Limiting abort level Grms on each
channel.
Password authorization to operate
the software.
Loop check feature thru Pre-test to
verify physical setting of the system
before starting actual test.
Independent alarm and abort profile
in terms of ± (1.0 to 50.0 %) of
reference pulse acceleration.
Auto test shutdown in case of control
exceeding the abort limit or lines
exceeds for set no. of loops.
Limiting abort level Grms on each
channel.
Password authorization to operate the
software.
Loop check feature thru Pre-test to
verify physical setting of the
system before starting actual test.
Independent alarm and abort
profile in terms of ± (1.0 to 50.0
%) of reference pulse acceleration.
Auto test shutdown in case of
control exceeding the abort limit or
lines exceeds for set no. of loops.
Limiting abort level Grms on each
channel.
Password authorization to operate
the software.
       
 
   Random-on-Random Control Software
Parameters Model - I Model - II Model - III
Independent no. of narrowband profile —- 1 to 10 1 to 10
Drive Signal —- Swept narrowband Gaussian
random combined with broadband
Gaussian random
Swept narrowband Gaussian random
combined with broadband Gaussian random
Breakpoint table —- Constant/ variable Constant/ variable
Sweep Rate —- Oct/min or Hz/sec or time/sec Oct/min or Hz/sec or time/sec
Star frequency —- Any value between the available
frequency range
Any value between
the available frequency range
Start direction —- Up/down Up/down
Run Schedule —- Broadband random on/off,
Narrowband sweeping on/off
Narrowband on/off
Broadband random on/off,
Narrowband sweeping on/off
Narrowband on/off
Safety —- User configurable ramp up rate in dB/sec User configurable ramp up rate in dB/sec
   Sine-on-Random Control Software
Independent no. of Sine tones 1 to 8 1 to 10 1 to 10
Drive Signal Multiple analog sweeping sine tone
combine with continuous broadband
Gaussian random
Multiple analog sweeping sine tone
combine with continuous
broadband Gaussian random
Multiple analog sweeping sine tone
combine with continuous
broadband Gaussian random
COLA output for Sine tone —- 1 1
Sweeping Fixed sine tones Fixed/ sweeping Fixed/ sweeping
Sweep Rate —- Oct/min or Hz/sec or time/sweep Oct/min or Hz/sec or time/sweep
Stat frequency Any value between the available
frequency range
Any value between the available
frequency range
Any value between the available
frequency range
Start direction —- Up/down Up/down
Sine tone level Constant Acceleration Constant Acceleration Constant Acceleration
Run Schedule Variable pre-test level & duration Variable pre-test level & duration
Broadband random on/off
Sine on/off per tone
Start sweep per tone
Variable pre-test level & duration
Broadband random on/off
Sine on/off per tone
Start sweep per tone
Safety Alarm/Abort limits

Alarm/Abort limits
User configurable ramp up rate in
dB/sec

Alarm/Abort limits
User configurable ramp up rate in
dB/sec
Storage of Test Data To disk at pre-configured intervals in
tabular text format & waveforms
To disk at pre-configured intervals in
tabular text format & waveforms
To disk at pre-configured intervals
in tabular text format & waveforms
Storage of Test Graph Storage of waveform into user
defined file.
Export of test graph to Microsoft
Office
Storage of waveform into user
defined file.
Export of test graph to Microsoft
Office
Storage of waveform into user
defined file.
Export of test graph to Microsoft
Office
Posttest Report Generation Facilitates report generation of test
conducted.
Export of graph to Microsoft Office
Facilitates report generation of test
conducted.
Export of graph to Microsoft Office
Facilitates report generation of test
conducted.
Export of graph to Microsoft Office
Safety features/
Protections
Loop check feature thru Pre-test to
verify physical setting of the system
before starting actual test.
Independent alarm and abort profile
in terms of ± (1.0 to 50.0 %) of
reference pulse acceleration.
Auto test shutdown in case of
control exceeding the abort limit or
lines exceeds for set no. of loops.
Limiting abort level Grms on each
channel.
Password authorization to operate
the software.
Loop check feature thru Pre-test to
verify physical setting of the system
before starting actual test.
Independent alarm and abort profile
in terms of ± (1.0 to 50.0 %) of
reference pulse acceleration.
Auto test shutdown in case of control
exceeding the abort limit or lines
exceeds for set no. of loops.
Limiting abort level Grms on each
channel.
Password authorization to operate the
software.
Loop check feature thru Pre-test to
verify physical setting of the
system before starting actual test.
Independent alarm and abort
profile in terms of ± (1.0 to 50.0
%) of reference pulse acceleration.
Auto test shutdown in case of
control exceeding the abort limit or
lines exceeds for set no. of loops.
Limiting abort level Grms on each
channel.
Password authorization to operate
the software.
       